Issue: 1.0
Routine Checks
Thermo Fisher Scientific
SAM12 User Manual 8-1
Chapter 8
Routine Checks
This section describes the routine checks required to ensure the
correct operation of the SAM12. Most mechanical and
electronic failures, if they occur, will become apparent during
normal operation and do not require checking. Therefore, they
are not included in this section.
Mechanical Checks
Once a month, or during source checks, check the operation of
the door lock, if the option is fitted. To do this, perform a
measurement with a source known to give alarms. When the
alarm occurs, check the door is locked and remains locked,
after the Residual Contamination Check if the option is
selected. Enter a valid password in the normal way (see
Gaining Access to the Administration Mode (page 5-3)
) and
check the door unlocks. For the two door SAM12 with door
locks, opening one door will lock the other.
Alternatively the locks can be exercised individually (see
Vault
(page 5-18)
for more information).
During calibration, or more frequently if required, check any
mechanical structure on which the SAM12 may be mounted.
Since the SAM12 with especially with 2" of lead shielding is
very heavy, any mounting arrangement should be inspected for
signs of deterioration, instability or any other factor which may
affect safety of operation or maintenance. If any such defect is
suspected, the SAM12 should be withdrawn from service and
safely removed using a suitable forklift and the integral fork
lifting facilities provided. The mounting should be repaired
and made safe before replacing the SAM12.
See
for details regarding
lubrication of the knuckles of the door hinge.
Any other mechanical faults should, if they occur, become
apparent during normal monitoring operation.
Door Locking Mechanism
(if fitted)
SAM12 Mounting
Arrangement
Other Mechanical Checks
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......