Specification
Issue: 1.0
3-4 SAM12 User Manual
Thermo Fisher Scientific
Programmable Threshold 3
0 - 4095mV in 1 mV steps
Programmable Threshold 4
0 - 4095mV in 1 mV steps
Programmable Threshold 5
0 - 4095mV in 1 mV steps
Control
Instrument ON/OFF operation is via a momentary key-switch
on the side panel. Operating any door activates a micro switch
and suspends background monitoring. Closing the door
indicates a sample is ready to be monitored and initiates a time-
out period. If the Start button is pressed within the time-out
period, the article is monitored, otherwise background
monitoring is resumed. Opening any door during monitoring
aborts the measurement. Optional door locks are available and
will activate to prevent any contaminated item being removed
from the clean door. Data entry and diagnostic functions are
accessed through a touch screen but access is password
protected.
Displays
The integral back-lit LCD display provided is used in
conjunction with the touch-screen for data entry and diagnostic
functions. As well as duplicating any front panel LED display,
the LCD can also provides comprehensive, user-friendly
operational guidance, measurement results and instrument fault
messages.
Five, bright, rear (optional front) panel lamps show the
instrument status at all times and give the user clear indication
of the measurement result.
FRONT BACK
(2-door
only)
Green:
Ready / Clear
Ready / Clear
White: Recount
Recount
Yellow: Count
Count
Red: Alarm
Alarm
Blue:
Out of Service
Out of Service
A pulsing white LED on the side panel indicating that the
mains is on and charging the battery.
Audible Indications
“Chime”
Measurement with a “CLEAR” result
“Single Beep”
End of a parameter counting sequence.
LCD
Lamps
Charging LED
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......