Issue: 1.0
Operating Instructions
Thermo Fisher Scientific
SAM12 User Manual 5-17
alarm conditions and background measurement - see
.
Select Diagnostics|Timed Counts
to view Detectors averaged over
a selected specified period. This
function allows a count to be
made which can be used to
ascertain the efficiency and
functionality of the SAM12
To toggle between the individual Detector's Thresholds and
the Gross Sum of all Detectors views
, press the
Detectors|Gross Sum button.
To Start the Timed Counts for the Detectors
:
1.
Select the number of SAM12 to be averaged by
selecting a value in the Average Over field.
2.
If you require a Continuous count to be made, tick the
Continuous box.
3.
Press Start button
To toggle between the Thresholds and Windows display,
tick the relevant radio button.
The variance check is a method by which the overall stability
of the background can be assessed by running a sequence of
background measurements and then assessing the variance of
the measured backgrounds over the sequence.
Select Diagnostics|Variance Test
to setup the background
monitoring over a number of
cycles to obtain a figure of
stability
Description of options:
Field Name
Description
Number of Cycles
Enter the number of background measurements to
be used in the test.
Cycle Duration
(seconds)
Enter the duration of each measurement.
Upper Variance Limit Enter the Upper Variance Limit for the result.
Timed Counts
Variance Test
Summary of Contents for SAM12
Page 1: ...MANUAL SAM12 SMALL ARTICLES MONITOR Thermo Fisher Scientific SAM12 21st September 2007...
Page 2: ......
Page 4: ...This manual was produced using ComponentOne Doc To Help...
Page 10: ......
Page 18: ......
Page 20: ......
Page 34: ......
Page 74: ...Operating Instructions Issue 1 0 5 34 SAM12 User Manual Thermo Fisher Scientific...
Page 130: ......
Page 134: ......
Page 144: ......
Page 168: ......