The SAM12 measures the
background and the calculated
count time decrements
If the calculated efficiency has an illegal value (i.e. no net
counts), the efficiency will be set to zero.
If a result is calculated, the new
calibration data is displayed.
Note that during a Cal Check
only an OK button is displayed
which returns to the
Calibration|Cal Check screen.
Note that the Cancel button will abandon the process.
Press the details button to reveal
the detailed results.
If you are satisfied with the
result, press the Confirm button
to display the Add Description
for the Calibration page. Select
OK to complete the process.
Note that the results are stored in
the database and marked as
PASS.
If you are NOT satisfied with the result, press the Fail button.
The Calibration Source page is re-displayed.
Note that the results are stored in the database and marked as
FAIL.