IMACC Script Setting
Modbus Control Version
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
E-13
For each system mode, the corresponding system control is configured by
the third script setting. The description for each system mode is
summarized in
Table B–3
in Appendix B, “
Modbus Protocol
”.
Each mode involves a couple of valve operations. Follow the similar
procedure as described in the section of CEMS_samplemode to set the
valve control setting. For instance, to configure the blowback valve to open,
just double click its setting block and set the correct DIO channel
“dev2/port1/line0:7”, pattern “0x050” and time delay setting “0”.
●
Script setting for ModbusMultiSensor
Next to the script setting for system command and status is the container
for “ModbusMultiSensor” setting, as shown.
Summary of Contents for Omni FTIR Multi Gas CEMS
Page 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 182: ......
Page 206: ......
Page 216: ......
Page 234: ......