Preventive Maintenance and Servicing
Pneumatic Panel Components
5-90
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
Use the following procedures to replace the pneumatic panel components.
Figure 5–62.
Pneumatic Panel Components Replacement
Use the following procedure to replace the pressure reducers on the
pneumatic panel. The procedure is the same for either pressure reducer.
Equipment Required:
Wrench, 9/16-inch
Adjustable wrench with a 1 1/8-inch minimum opening
Pressure reducer
1.
Remove the left side panel per procedure “
Removing the Left Side
Panel
” for better accessibility.
2.
Note the pressure on the gauge above the pressure reducer to be
replaced, so that it can be reset later.
Pneumatic Panel
Components
Replacing the
Pressure Reducers