Preventive Maintenance and Servicing
Maintenance and Servicing of the FTIR Analyzer Components
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
5-11
6.
Using the 5/16-inch hex wrench, install the new pre-filter into the
stinger and tighten until snug.
7.
Re-install the probe tip shield onto the end of the stinger. Using the
3/8-inch wrench and the 5/32-inch hex wrench, tighten the locknut
and screw on the probe tip shield until snug.
8.
Remove the flange cap from the stack opening. Carefully re-install the
probe into the stack, making sure the gasket is properly seated. Thread
the 4 flange nuts and washers onto the threaded studs, and then tighten
using the 15/16-inch wrench until snug.
9.
Power up the probe by turning on the Probe circuit breaker on the
breaker box to complete the procedure.
Use the following procedures to perform maintenance and servicing on the
FTIR analyzer. Maintenance procedures are performed periodically to keep
the analyzer running efficiently. Service procedures are performed to
replace a failing component in the analyzer.
WARNING
Perform only those procedures on the analyzer described in this
section. If there are other problems with the analyzer, contact technical
support. Also refer to the Antaris IGS manual in the “
Other Manuals
” on
page
1-10
.
▲
Use the following procedure to periodically check and/or replace the
desiccant packet in the FTIR analyzer, which is located behind the front
panel. These packets keep the analyzer optics free of excess moisture. They
should be checked even if the analyzer is in storage. Replace spent desiccant
packets, as they cannot be dried or regenerated safely. Keep all covers closed
tightly unless a maintenance and servicing procedure is being performed, to
minimize the exposure to humidity.
WARNING
The desiccant is harmful if swallowed. Follow the
manufacturer’s instructions to dispose of desiccant packets properly.
▲
CAUTION
Avoid burn and fire hazards. Do not attempt to regenerate spent
desiccant packets.
▲
Maintenance and
Servicing of the
FTIR Analyzer
Components
Checking and
Replacing the
Desiccant
Summary of Contents for Omni FTIR Multi Gas CEMS
Page 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 182: ......
Page 206: ......
Page 216: ......
Page 234: ......