Troubleshooting
Boot Up Error/IMACC Software Error
6-4
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
●
Bad spectrometer detector or source
●
Check ‘Last Vmax’ and ‘Last Vmin’ in ‘Status’. Nominally Last
Vmax >4.0 Volts, Last Vmin <-1.0 Volts. Readings of <±0.5
Volts indicate detector or source failure. Marginal readings may
require more detailed analysis by Thermo Fisher Scientific
Technical Support Department.
●
Concentration not updating in IMACC
●
See ‘Communications Errors/No Communications with FTIR’.
●
H
2
O concentrations <2000 ppm during zero checks
●
Make sure hydrator is not filled with water.
System does not start automatically
●
Verify the following programs are in the setup:
Omni.layout
ImaccCurrentLoop
ImaccModbus
ImaccModbusTCP
●
See “
System Startup
” in Chapter 2.
Bad temperature readings
– Either in ‘Status’ or in Modbus block in
IMACC script (See Appendix E, “
IMACC Script Setting
”.)
●
“Last Temperature” reading in ‘Status’ on ImaccMonitor software
main screen reads 185.000000.
●
Check virtual com port number (See Appendix D, “
Calogix
Setup
”).
●
Check USB connection between PC and temperature
controller.
●
Stop and re-start script in ImaccMonitor.
●
“Pump” sampling box temperature reading is low, but other
temperatures are OK.
●
Check thermostat inside pump box (See Chapter 5, “
Preventive
Maintenance and Servicing
”.
●
ERROR_QUANT_TEMPRESSCON in ImaccMonitor
●
Check temp control breaker and AC wiring connections.
Boot Up
Error/IMACC
Software Error
Temperature
Errors
Summary of Contents for Omni FTIR Multi Gas CEMS
Page 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 182: ......
Page 206: ......
Page 216: ......
Page 234: ......