Preventive Maintenance and Servicing
Replacing the Gas Cell Heater
5-52
Omni FTIR Multi Gas CEMS Instruction Manual
Thermo Fisher Scientific
Gloves
Flat blade screwdriver
High temperature tape
Gas cell heater
CAUTION
High surface temperatures present. Allow sufficient time to
cool.
▲
1.
Shut down the FTIR heaters by turning off the Temperature Control
circuit breaker on the breaker box and allow it to cool for 1 hour. This
will put the system in fail safe mode so that it will not be sampling
stack gas.
2.
Remove the gas cell by following procedure “
Replacing the Gas Cell
”.
3.
Using the utility knife, carefully cut the paper of the insulation at the
seam (
Figure 5–42
).
Figure 5–42.
Cutting Insulation Paper at the Seam
4.
Remove the insulation from the gas cell. Wearing gloves is suggested
when handling the insulation (
Figure 5–43
).