Concentration Data Transfer Using a Web Server
Modbus Control Version
Thermo Fisher Scientific
Omni FTIR Multi Gas CEMS Instruction Manual
F-3
Test of Concentration Data Retrieval
1.
To test the retrieval process, copy any concentration data file on the
FTIR computer to the data transfer folder and rename it
current.con
.
2.
Open Internet Explorer on the remote computer.
3.
In the IE Address bar enter
http://<address>/FTIR Data/current.con
,
where <
address
> is either the FTIR computer’s name on the network or
its IP address.
4.
Internet Explorer will return the concentration file.
5.
The extraction and use of the data returned by IE is the responsibility
of the individual user. However, the use of IE for data transfer is not
necessarily required; users may develop their own IIS-compatible
programs to access and transfer the data.
Configuration of IFSS for IIS Data Transfer (FTIR Computer)
1.
Open
IMACC Script Editor
and open the data collection script.
2.
Double-click on the
CollectEZ
window.
3.
Click on the options tab.
4.
In the entry box
Copy CON file to directory,
browse for the data transfer
folder created earlier. IFSS will save a copy of the current concentration
file to the file
current.con
in that folder.
5.
Click
OK
and save the script.
Summary of Contents for Omni FTIR Multi Gas CEMS
Page 1: ...Omni FTIR Multi Gas CEMS Instruction Manual Part Number 108004 00 11Apr2013...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 20: ...Figures xvi Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 22: ...Tables xviii Omni FTIR Multi Gas CEMS Instruction Manual Thermo Fisher Scientific...
Page 182: ......
Page 206: ......
Page 216: ......
Page 234: ......