The Specs Screen
Thermo Scientific
NITON XL3 Analyzer User’s Guide
1-71
Bias:
Bias should be within these ranges:
Standard 6 mm Detector:
175 + 10
GOLDD SDD Detector:
-220 + 15
VCool:
VCool will vary with the ambient temperature.
SubBias:
SubBias should be within these ranges:
Standard 6 mm Detector:
-11 + 3
GOLDD SDD Detector:
-8 + 3
Escale:
Escale should be within these ranges:
Standard 6 mm Detector or
GOLDD SDD Detector:
6.6 - 9
Preamp:
Preamp value should only be noted, and reported to Service if there is a
problem.