To Prepare or Not to Prepare - In Situ vs. Ex Situ
Thermo Scientific
NITON XL3 Analyzer User’s Guide
1-35
To Prepare or Not to
Prepare - In Situ vs.
Ex Situ
Bulk media are generally tested either on-site (
in situ
) for screening
purposes, or removed and prepared (
ex situ
) to enhance the accuracy of the
measurement. The degree of preparation may vary depending on the
accuracy desired, the characteristics of the sample, and the characteristics of
the site.
Understanding the advantages of
in situ
testing and of various degrees of
preparation is crucial to obtaining useful data.
In situ
measurements should
be used to profile an area, to locate areas of interest, to determine the
boundaries of such areas, or to gather data that will subsequently be used to
design a sampling plan.
In situ
measurements are usually only approximations, though they may
correlate very well with lab analysis if the site tested is highly homogeneous.
If the site is non-homogeneous, as is often the case, then in situ
measurement results may differ greatly from laboratory obtained results.
Both sets of results may be correct. The difference arises from the fact that
actual samples tested were different.
Summary of Contents for NITON XL3t 500
Page 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Page 2: ......
Page 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...