Mining Ta/Hf Mode
1-32
NITON XL3 Analyzer User’s Guide
Thermo Scientific
Concentrations for the following analytes can be determined:
Ti, V, Cr, Mn, Fe, Co, Ni, Hf, Re, W, Pb, Bi, Zr, Ta, Nb, Mo, Sn, Ba, Sb,
Cd, Pd, Sr, Rb, Se, and Ag.
All concentrations are displayed in units of wt. % by default, but can be
changed to ppm.
Note
For defensible Quality Control, keep a record of the time and
precision of every calibration
.
WARNING!
Tampering with the 5,500 ppm (Lead high) lead-in-soil
standard may cause exposure to lead dust. Keep all standards out of reach of
children.
CAUTION
Never tamper with Test Standards. They should not be used
unless they are completely intact
.
During each test, the instrument looks at the full range of x-ray spectrum
and continuously corrects for cross-element interference.
Testing Prepared Samples
Set the NITON test platform on a flat, solid surface. Open the lid and place
the sample cup in the holder, then shut the lid. Insert the instrument into
the nose cone adaptor and follow in-situ bulk sample instructions
Summary of Contents for NITON XL3t 500
Page 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Page 2: ......
Page 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...