The Measurement Screen
Thermo Scientific
NITON XL3 Analyzer User’s Guide
1-33
The Measurement
Screen
Figure 1-1. The Standard Bulk Measurement Screen
The Result Screen
When you end a reading, the
Measurement
screen will be replaced by the
Result
screen. These screens displays the reading grouped as follows:
At the top, elements detected in the sample, and, underneath this, elements
that were below the detection limit.
Use the 4 way touch pad to scroll through the elements.
Detection Limit
For an element to be detected by the analyzer in a given sample, the
measured concentration of the sample must be at least three times the
standard deviation of the measurement. This detection limit will depend on
the composition of the sample.
Precision
The measurement precision for each element displayed appears to the right
of the measured concentration, under the heading “
+-
“. The precision of
each measurement is two times the standard deviation (sigma).
Element
Column
NAV Menu
Two Sigma
Confidence
Column
Mode
Nominal Seconds
Test Duration Line
Reading Number
Tools Menu
Tools Menu
Tools Menu
Tools Menu
Battery Life
Indicator
Measured
Concentration
Summary of Contents for NITON XL3t 500
Page 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Page 2: ......
Page 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...