The Calibrate Touch Screen Screen
Thermo Scientific
NITON XL3 Analyzer User’s Guide
1-63
The Calibrate Touch
Screen Screen
Figure 1-20. The Calibrate Touch Screen Menu Path
Select the
Calibrate Touch Screen
icon to re-calibrate the analyzer's touch
screen display. This procedure establishes the display boundaries for the
touch screen interface. When the
Calibrate Touch Screen
icon is selected,
the display will show the message: "Calibrate Touch Screen". There will be a
small cross in the upper left-hand corner of the display. Tap on this cross
with the stylus, and the cross will disappear and reappear in the upper
right-hand corner of the screen. Tap on the cross again, and it will reappear
in the lower right-hand corner of the screen. Tap on the cross again and it
will reappear in the lower left-hand corner of the screen. Tap on the cross
once more, and you will be presented with the
Calibrate Menu