Analysis of Unprepared Samples – In Situ
Thermo Scientific
NITON XL3 Analyzer User’s Guide
1-41
Because you are testing through a bag, test results will tend to be lower
than test results obtained from direct analysis. This effect will vary
depending on the element analyzed and the thickness and composition of
the cloth through which the sample is tested. Bagged samples can be
retested and/or be further prepared and then retested, allowing samples of
particular interest to be more accurately analyzed.
Environmental
Environmental testing is used to assure that the site stays within the
environmental guidelines of the government overseeing the operation.
Environmental testing can give warning of possible environmental guideline
violations, giving the site ownership opportunity to fix the problem before
regulatory processes intervene. Areas downwind and downstream from the
site should be patricularlyclosely moitored for higher than normal levels of
metals.
Screening Techniques
Use direct measurement when you need to determine whether an element is
present (rather than in accurately measuring how much is present). Use
preliminary direct measurements to survey a site quickly even if you intend
to take samples
WARNING!
When taking samples from a site where toxic chemicals may be
present, always use gloves and respiration equipment for your own
protection.
1. Select a measurement site, and clear away any surface debris and
vegetation.
Note
Valid results will depend on a sufficient and appropriate selection of
sites to sample. Lead-in-soil from paint, for instance, will usually be
concentrated within a few feet of the painted structure.
2. Choose an area to test where the measurement window of the
analyzer will be flush with the test media. Position the nose
against the surface to be analyzed and initiate a reading by
squeezing the shutter release, and firmly pressing the instrument
flat against the surface.
3. After the test, inspect the nose of the instrument for
contamination, which may affect future analysis. If the nose
appears to be soiled, clean it with a soft cloth or tissue.
Summary of Contents for NITON XL3t 500
Page 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Page 2: ......
Page 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Page 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...