The Specs Screen
1-70
NITON XL3 Analyzer User’s Guide
Thermo Scientific
Diagnostics
Select the Diagnostics Button to load the Diagnostics Screen. the
Diagnostics Screen shows Detector Temperature, Bias, Cooler Voltage,
SubBias, Energy Scale, and Temperature in C and F scales.
The Diagnostics Screen can be of great utility in assuring proper operation
of your analyzer.
Figure 1-26. Diagnostics Screen
The proper ranges of operational values on the Diagnostics Screen follow.
Det Temp:
Detector Temperature should be within these ranges:
Standard 6 mm Detector:
-25 + 5 degrees F
GOLDD SDD Detector:
-27 + 3 degrees F