UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
152
Operating Instructions
Message
Remedial Action
Implausible left working
cylinder pressure. Check
pressure plug.
—or—
Implausible right working
cylinder pressure. Check
pressure plug.
The pressure values established during pump operation for the
working pressure of the left and/or right pump head of the NC pump
are not plausible.
The cable for the related pressure transducer may not be connected
properly. Verify that the cable properly connected.
Perform the following sequence of actions:
1.
ProFlow flow meter:
Perform the
Zero Balance Test
(
→
page 172).
Classic flow meter:
Run the
Pressure Transducer Test
(
→
page 172).
2. Purge the pump head for which the message appears (
→
page 73).
3. Close the flow meter outlet with a Viper fitting plug (included in
the accessories kit of the module) and verify that the purge valves
are closed.
4. From Chromeleon, perform the
CalibrateWP
command for the
pump head.
5. Purge the flow meter (
→
page 75).
Replace the pump head (
→
page 188).
Implausible left working
cylinder calibration. Check
pressure plug.
—or—
Implausible right working
cylinder calibration. Check
pressure plug.
The calibration values established during pump operation for the left
and/or right pump head of the NC pump are not plausible.
The cable for the related pressure transducer may not be connected
properly. Verify that the cable properly connected.
Perform the following sequence of actions:
1.
ProFlow flow meter:
Perform the
Zero Balance Test
(
→
page 172).
Classic flow meter:
Run the
Pressure Transducer Test
(
→
page 172).
2. Purge the pump head for which the message appears (
→
page 73).
3. Close the flow meter outlet with a Viper fitting plug (included in
the accessories kit of the module) and verify that the purge valves
are closed.
4. From Chromeleon, perform the
CalibrateWP
command for the
pump head.
5. Purge the flow meter (
→
page 75).
Replace the pump head (
→
page 188).
Internal pump maintenance is
due, please open purge screw
and purge
Internal maintenance of the loading pump is overdue and has not
been performed for more than 45 days. Open the purge screw and
initiate a purge cycle.
Leak detected
When this message appears, the flow is automatically turned off after
180 seconds.
The leak sensor in the pump module has reported a leak.
Find and eliminate the source for the leak. Dry the leak sensor and
tray (
→
page 177).
Left valve no head found.
While the Catch Head command has been performed, no valve has
been found. Repeat the command and push the valve onto the
actuator (
→
page 78).
Summary of Contents for Dionex UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Page 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Page 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Page 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Page 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Page 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...