UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Operating Instructions
Page
121
5.5.6 Recording the Pump Signals
On the
Signals
page, the check box for the column pressure of the NC pump and the
pressure of the loading pump are selected by default when the module is installed and
configured in Chromeleon (
→
page 51).
Depending on the flow meter that is installed, observe the following:
•
Make sure that the corresponding signals for the flow meters (see table below) are
selected to have these signal channels available for recording the pump flow (ProFlow
flow meter only) or pump pressure (Classic flow meter only) in Chromeleon.
•
If you exchanged a flow meter type (for example from Classic to ProFlow)
Make sure that the signals for the other flow meter type are deactivated (see the example
below).
When signals are selected, Chromeleon generates the appropriate channels for recording the
signals. The channels are then available in the
Commands
dialog box for the pump. The
channel name is generated from the entry in the
Pump Device Name
box on the
Pumps
page
(
→
page 48).
If a problem occurs, the signals can provide helpful information to identify and eliminate the
source for the problem. Therefore, always record the signals of your module.
The signal ... (default name)
Records the following signal ...
NC_Pump_Pressure
Column pressure of the NC pump.
Selected by default.
LoadingPump_Pressure
Pressure of the loading pump.
Selected by default.
ProFlow flow meter only
If a ProFlow flow meter is installed, make sure that these signals are
activated.
NC_Pump_Flow_LeftBlk
Flow rate of the left flow cannel (A) of the NC pump. The flow is
measured in nL/min.
NC_Pump_Flow_RightBlk
Flow rate of the right flow cannel (B) of the NC pump. The flow is
measured in nL/min.
NC_Pump_Flow
Total flow rate from the left and right flow channel flow rates
(
NC_Pump_Flow_LeftBlk
and
NC_Pump_Flow_RightBlk
).
Classic flow meter only
If a Classic flow meter is installed, make sure that these signals are
activated.
NC_Pump_Press_LeftBlk
The primary pressure of the left or right pump head of the NC pump. If
the pressure is above 100 MPa, pumping stops.
NC_Pump_Press_RightBlk
Example
If a you exchange a ProFlow flow meter with a Classic flow meter, make sure that the
NC_Pump_Flow
,
_LeftBlk
and
_RightBlk
signals are deactivated and the
NC_Pump_Press_LeftBlk
and
_RightBlk
signals for the Classic flow meter are activated on
the
Signals
page.
Summary of Contents for Dionex UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Page 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Page 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Page 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Page 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Page 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...