UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
92
Operating Instructions
5.3.3 Automated Control
With automated control, you create a program file (PGM) for automated operation of the
module. You can create programs automatically with the software wizard or manually by
editing an existing program. In addition to programs for sample analysis, you can also
create programs for special purposes, for example, to automate system shutdown
(
→
page 141) or to ensure that the system automatically restarts operation as desired after a
power failure. For details, see the
Chromeleon Help
.
To create a program with the Program Wizard
1.
Open the Program Wizard. On the
File
menu, select
New
, and then select
Program
File
.
2.
The wizard guides you through program creation. On each wizard page, accept or
change the settings as appropriate. For additional information about a page, click
Help
.
3.
After you finish the wizard, Chromeleon automatically creates the corresponding
program.
4.
To start the program, follow the steps below (
→
page 93).
To create a program manually
1.
Open an existing program.
Select and double-click the program you want to open.
—or—
On the
File
menu, select
Open
. In the dialog box, select
Program
on the
Object of Type
list and select the program.
Fig. 47: Chromeleon program (here program shown in the Commands view)
Open the
Device Views
Opens the
Commands View
Summary of Contents for Dionex UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Page 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Page 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Page 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Page 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Page 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...