UltiMate 3000 Series:
NCS-3500RS and NCP-3200RS
Page
42
Operating Instructions
3.4.3 Connecting the Power Cord
Use the power cord shipped with the module to connect the module to the main power
source. Connect the power cord from the main power receptacle on the rear panel
(
→
Fig. 2, page 24) to a grounded power source. No manual adjustment is required to
adapt the line voltage to local voltage requirements.
Warning:
Never use a power cord other than the power cords provided for the
device.
Do not use multiple sockets or extension cords. Using defective
multiple sockets or extension cords may cause personal injury or
damage to the device.
Avertissement:
Utilisez uniquement les cordons d’alimentation électrique spécifique
à l’instrument.
N'utilisez pas des blocs multiprise ou des câbles prolongateurs. Cela
pourrait entraîner des blessures corporelles ou endommager
l'instrument.
3.4.4 Connecting a Solvent Rack
If the UltiMate 3000 system includes a SRD-3x00 Solvent Rack, connect the
Solvent
Rack
port on the rear panel of the module to the 15-pin D-Sub connector on the rear panel
of the Solvent Rack. The appropriate connection cable is included in the accessories kit of
the Solvent Rack.
3.4.5 Connecting Devices (Mass Spectrometers) to the Digital I/O-Port
To connect a device, for example, a mass spectrometer, to a digital I/O port on the rear
panel of the module, use a 6-pin mini-DIN signal cable (part no. 6000.1004). The cable is
available from Thermo Fisher Scientific.
1.
Plug the 6-pin connector of the mini-DIN cable into the
Digital I/O
port
1
(or
2
). For
information about the functions of the connector pins and pin assignment, see
page 255.
2.
For each relay output or digital input to be used, connect the appropriate signal wire
and ground wire to the corresponding connectors on the device (
→
documentation
provided with the device).
For example, if you want to use a relay output to control a mass spectrometer, use the
wires 5 (brown) and 6 (white).
3.
When setting up the module in the Chromeleon
Server Configuration
program
(
→
page 46), select the corresponding relay outputs on the
Relays
page and the digital
inputs on the
Inputs
page.
The relay outputs and digital inputs are then available in Chromeleon, for example, in the
Commands
dialog box under
PumpModule
, and can be programmed as required.
Summary of Contents for Dionex UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Operating Instructions ...
Page 8: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 16 Operating Instructions ...
Page 62: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 54 Operating Instructions ...
Page 92: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 84 Operating Instructions ...
Page 246: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 238 Operating Instructions ...
Page 262: ...UltiMate 3000 Series NCS 3500RS and NCP 3200RS Page 254 Operating Instructions ...