The
Pressure Transducer Test
must
be performed when the following components have
been replaced:
♦
Pump head
♦
Flow meter
♦
Flow selector
•
As long as the retention times are stable and in the expected range, you should
not
interrupt pump operation by recalibrating the pressure transducers.
•
You should check also the offset of the pressure transducers:
♦
Before
you test the NC pump for leakage by running the diagnostics functions from
Chromeleon. Observe the recommended order of tests (
→
page 134).
♦
If you observe a shift in peaks or if the gradients are not correct.