Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-51
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config >
All Voltage Channels, All Current
Channels, Voltage Channel 1–6 or Current Channel 1–6
.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE AVG DIAGS ALARM
The Select Range screen is used to select the hardware range for the selected
analog output channel. Possible ranges for the voltage outputs are: 0-100
mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs are: 0-20
mA and 4-20 mA.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > select Channel >
Select Range
.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-100mV ?
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Minimum and Maximum Value screens are used to edit the zero (0)
and full-scale (100) values, respectively, in percentages for the selected
analog output channel. See
Table 3–6
for a list of choices. In dual or
autorange mode, “HI” or “LO” is displayed to indicate high or low
concentrations. Range Status is visible only in auto range mode. The O
2
conc items are only displayed if the internal O
2
sensor option is installed.
The minimum and maximum output value screens function the same way.
The following example shows the set minimum value screen.
●
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > select Channel >
Set Minimum
or
Maximum
Value
.
Select Range
Minimum and Maximum
Value