C-Link Protocol Commands
Alarms
B-18
Model 410
i
Instruction Manual
Thermo Fisher Scientific
set alarm sample flow min
value
set alarm sample flow max
value
These commands set the sample flow alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing sample
flow alarm limits in liters per minute. The following example sets the
sample flow alarm maximum value to 1 L/min.
Send:
set alarm sample flow max 1.000
Receive:
set alarm sample flow max 1.000 ok
alarm trig conc
This command reports the CO
2
concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The following
example shows the CO
2
concentration minimum alarm trigger to ceiling,
according to
Table B–4
.
Send:
alarm trig conc
Receive:
alarm trig conc 1
set alarm trig conc
value
This command sets the CO
2
concentration alarm minimum
value
, where
value
is set to either floor or ceiling, according to
Table B–4
. The following
example sets the CO
2
concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc 1
Receive:
set alarm trig conc 1 ok
Table B–4.
Alarm Trigger Values
Value
Alarm Trigger
00 Floor
01 Ceiling
alarm trig conc o2
This command reports the O
2
concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The following
example shows the O
2
concentration minimum alarm trigger set to ceiling,
according to
Table B–4
.
Send:
alarm
trig
conc
o2
Receive:
alarm trig conc o2 1