C-Link Protocol Commands
Alarms
B-14
Model 410
i
Instruction Manual
Thermo Fisher Scientific
Figure B–1.
Flags Field
alarm agc intensity min
alarm agc intensity max
These commands report the agc intensity alarm minimum and maximum
value current settings. The following example reports that the agc intensity
alarm minimum value is 20.
Send:
alarm agc intensity min
Receive:
alarm agc intensity min 20
set alarm agc intensity min
value
set alarm agc intensity max
value
These commands set the agc intensity alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing agc
intensity alarm limits. The following example sets the agc intensity alarm
maximum value to 20.
Send:
set alarm agc intensity max 20
Receive:
set alarm agc intensity max 20 ok
alarm bias voltage min
alarm bias voltage max
Alarms