C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-27
stream time
This command reports if the streaming data string will have a time stamp
attached to it or not, according to
Table B–6
. The following example
reports that the streaming data shall not include a time stamp.
Send:
stream
time
Receive:
stream time 0
set stream time
value
This command enables
value
, where
value
is to attach or disable time stamp
to streaming data string, according to
Table B–6
. The following example
attaches a time stamp to streaming data.
Send:
set stream time 0
Receive:
set stream time 0 ok
Table B–6.
Stream Time Values
Value
Stream Time
0
Disables time stamp to streaming data string
1
Attaches time stamp to streaming data string
set cal co2 bkg
set zero calibrate
These commands will set the CO
2
background. The following example
shows a successful auto-calibration of the CO
2
background.
Note
Zero air must be provided to the instrument prior to issuing any of
these commands.
▲
Send:
set cal co2 bkg
Receive:
set cal co2 bkg ok
cal bkg o2
This command will set O
2
background so the O
2
concentration reads
0.00%. The example below shows a successful auto-calibration of the O
2
background.
Send:
set cal bkg o2
Receive:
set cal bkg o2 ok
Calibration