Operation
Instrument Controls Menu
3-44
Model 410
i
Instruction Manual
Thermo Fisher Scientific
The IP Address screen is used to edit the IP address. The IP address can
only be changed when DHCP is off. If DHCP is on, the instrument will
respond with “NOT SETTABLE IF DHCP IS ON”. For more
information on DHCP, see “
Use DHCP
” above.
●
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings
>
IP Address
.
IP ADDRESS:
CURRENT: 192.168.1.200
SET TO: 192.168.1.200 1
MOVE CURSOR
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Netmask screen is used to edit the netmask. The netmask is used to
determine the subnet on which the instrument can directly communicate
to other devices. The netmask can only be changed when DHCP is off. If
DHCP is on, the instrument will respond with “NOT SETTABLE IF
DHCP IS ON”. For more information on DHCP, see “
Use DHCP
”
above.
●
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings
>
Netmask
.
NETMASK:
CURRENT: 255.255.255.0
SET TO: 255.255.255.0 0
MOVE CURSOR
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Default Gateway screen is used to edit the gateway address. The
default gateway can only be changed when DHCP is off. If DHCP is on,
the instrument will respond with “NOT SETTABLE IF DHCP IS ON”.
For more information on DHCP, see “
Use DHCP
” above. Any traffic to
addresses that are not on the local subnet will be routed through this
address.
●
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings
>
Gateway
.
IP Address
Netmask
Default Gateway