![Thermo Scientific 410i Instruction Manual Download Page 243](http://html1.mh-extra.com/html/thermo-scientific/410i/410i_instruction-manual_1104586243.webp)
C-Link Protocol Commands
Commands
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-3
Table B–1.
Command Response Error Descriptions
Command Response
Description
bad cmd
Command is not recognized
too high
Supplied value is higher than the upper limit
too low
Supplied value is lower than the lower limit
invalid string
Supplied string invalid (typically because a letter was detected
when the value should be numeric)
data not valid
Supplied value is not acceptable for entered command
can’t, wrong settings
Command not allowed for current measurement mode
can’t, mode is service
Command not allowed while instrument is in service mode
The “save” and “set save params” commands (duplicated for backward
compatibility) store parameters in FLASH memory. It is important that
this command be sent each time instrument parameters are changed. If
changes are not saved, they will be lost in the event of a power failure.
When interfacing to an instrument via C-link commands, always enter the
concentration values in ppb or μg/m
3
units. For example, to set a
background value to 20 ppm, enter 20000 (ppb) as the value for the set
background command.
Convert concentration formats from xxxxE+yy to x.xxxE+yy. The
i
Series
uses more standard script than the C-Series by placing only one digit to the
left of the decimal.
Table B–2
lists the 410
i
C-Link protocol commands. The interface will
respond to the command strings outlined below.
Table B–2.
C-Link Protocol Commands
Command
Description
Page
1
Simulates pressing soft key 1 pushbutton
B-34
2
Simulates pressing soft key 2 pushbutton
B-34
3
Simulates pressing soft key 3 pushbutton
B-34
4
Simulates pressing soft key 4 pushbutton
B-34
addr dns
Reports/sets domain name server address for Ethernet port
B-41
addr gw
Reports/sets default gateway address for Ethernet port
B-42
addr ip
Reports/sets IP address for Ethernet port
B-42
Entering Units in
PPB
Convert Concentration
Formats
Commands List