C-Link Protocol Commands
Calibration
B-32
Model 410
i
Instruction Manual
Thermo Fisher Scientific
span dur
This command reports the span duration. The following example reports
that the span duration min is 10 minutes.
Send:
span
dur
Receive:
span dur 10 min
set span dur
value
This command sets span duration to
value
where
value
represents the span
duration in minutes. The following example sets the span duration to 15
minutes.
Send:
set span dur 15
Receive:
set span dur 15 ok
zero cal reset
This command reports that the zero cal reset is on or off. The following
example reports that the zero cal reset is off.
Send:
zero cal reset
Receive:
zero cal reset off
set zero cal reset
onoff
onoff
= | on | off |
This command sets the zero cal reset on or off. The following example
turns the zero cal reset off.
Send:
set zero cal reset off
Receive:
set zero cal reset off ok
zero dev
This command reports the zero deviation (maximum zero check offset).
The following example reports that the zero deviation is 10 ppb.
Send:
zero
dev
Receive:
zero dev 1.000 E+01
set zero dev
value
This command sets the zero deviation (maximum zero check offset) to
value
, where
value
is a floating-point representation of the gas
concentration in current selected units. The following example sets the zero
deviation to 10 ppb.
Send:
set zero dev 1.000 E+01
Receive:
set zero dev 1.000 E+01 ok