C-Link Protocol Commands
Measurements
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-11
Selection
Averaging Time (seconds)
7 90
8 120
9 180
10 240
11 300
co2
high co2
low co2
These commands report the measured CO
2
concentration when operating
in single range, or high and low CO
2
when operating in dual or auto range
mode. The following example shows that the CO
2
concentration is 40
ppm.
Send:
CO2
Receive: CO2
4.000E+1
ppm
o2
This command reports the oxygen concentration in percent. The following
example reports that the current O
2
concentration is 15%.
Send:
o2
Receive:
o2 15.00 %
o2 corr conc
This command reports the oxygen correction concentration in percent.
The following example shows that the O
2
correction concentration is 15%.
Send:
o2 corr conc
Receive:
o2 corr conc 15.00 %
set o2 corr conc
value
value
= 0 to 20.9
This command sets the oxygen correction concentration value to
value
,
where
value
is a floating-point number representing the oxygen correction
concentration in percent. The following example sets the O
2
corrected
concentration value to 15.00%.
Send:
set o2 corr conc 15
Receive:
set o2 corr conc 15 ok