C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-15
These commands report the bias voltage alarm minimum and maximum
value current settings. The following example reports that the bias voltage
alarm minimum value is -128.
Send:
alarm bias voltage min
Receive:
alarm bias voltage min -128
set alarm bias voltage min
value
set alarm bias voltage max
value
These commands set the bias voltage alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing bias
voltage alarm limits. The following example sets the bias voltage alarm
maximum value to -110.
Send:
set alarm bias voltage max -110
Receive:
set alarm bias voltage max -110 ok
alarm chamber temp min
alarm chamber temp max
These commands report the bench temperature alarm minimum and
maximum value current settings. The following example reports that the
bench temperature alarm minimum value is 35.0 °C.
Send:
alarm chamber temp min
Receive:
alarm chamber temp min 35.0 deg C
set alarm chamber temp min
value
set alarm chamber temp max
value
These commands set the bench temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing bench temperature alarm limits in degrees C. The following
example sets the bench temperature alarm maximum value to 55.0 °C.
Send:
set alarm chamber temp max 55.0
Receive:
set alarm chamber temp max 55.0 ok
alarm conc min
alarm conc max
These commands report the CO
2
concentration alarm minimum and
maximum values current setting. The following example reports that the
CO
2
concentration minimum is 5.2 ppm.
Send:
alarm conc min
Receive:
alarm conc min 5.2 ppm
set alarm conc min
value
set alarm conc max
value