C-Link Protocol Commands
I/O Configuration
B-52
Model 410
i
Instruction Manual
Thermo Fisher Scientific
set dout
channel index state
This command assigns digital output
channel
(1-10) to be assigned to the
action associated with
index
(1-34), and assigns it an active state of
state
(open or closed). Use the “list var dout” command to obtain the list of
supported index values and corresponding state. The following example sets
the digital output channel 4 to state 11.
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa
channel
This command reports the outputs of the 6 or 12 Digital to Analog
converters, according to
Table B–15
. The following example shows that the
D/A #1 is 97.7% full-scale.
Send:
dtoa
1
Receive: dtoa
1
97.7%
Note
If the instrument is in a mode which does not provide a particular
output, and that output is selected, the value will be 0.0.
▲
Note
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
▲
Table B–15.
Default Analog Output Channel Assignments
D to A
Function
Single Range
Dual Range
Auto Range
1 Voltage
Output
CO
2
Low
CO
2
CO
2
2 Voltage
Output
CO
2
High
CO
2
Range
Status
3
Voltage Output
Not Used
Not Used
Not Used
4
Voltage Output
Not Used
Not Used
Not Used
5
Voltage Output
Not Used
Not Used
Not Used
6
Voltage Output
Not Used
Not Used
Not Used
7 Current
Output
CO
2
Low
CO
2
CO
2
8 Current
Output
CO
2
High
CO
2
Range
Status
9
Current Output
Not Used
Not Used
Not Used
10
Current Output
Not Used
Not Used
Not Used
11
Current Output
Not Used
Not Used
Not Used
12
Current Output
Not Used
Not Used
Not Used