Servicing
Filter Wheel Replacement
Thermo Fisher Scientific
Model 410
i
Instruction Manual
7-13
Cover Screws (2)
IR Source Mounting Posts (2)
Motor Plate
IR Source
Figure 7–6.
Replacing the IR Source
Use the following procedure to replace the filter wheel.
Equipment Required:
Filter wheel
Allen wrenches, 5/32-inch and 5/64-inch
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Filter Wheel
Replacement
IR Source Cover
Clamp Screws (2)
Filter Wheel Set Screw
Filter Wheel Mount
Access Hole
Filter Wheel