![Thermo Scientific 410i Instruction Manual Download Page 57](http://html1.mh-extra.com/html/thermo-scientific/410i/410i_instruction-manual_1104586057.webp)
Operation
Calibration Menu
Thermo Fisher Scientific
Model 410
i
Instruction Manual
3-21
CAL O2 COEFFICIENT
The Calibrate CO
2
Background screen is used to adjust the instrument
zero, or background. Before making an adjustment, be sure the analyzer
samples zero air for at least 5 minutes.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “
Calibration
”.
●
In the Main Menu, choose Calibration >
Cal CO2 Background
.
CO2 BACKGROUND:
CO2: 15.8
CURRENTLY: 0.0 ?
SET CO2 TO ZERO
RANGE AVG DIAGS ALARM
The Calibrate CO
2
Coefficient screen is used to adjust the CO
2
coefficient
and enter the span concentration. The display shows the current CO
2
concentration reading and the current CO
2
range. The next line of the
display is where the CO
2
calibration gas concentration is entered.
The CO
2
span coefficient is calculated, stored, and used to correct the
current CO
2
reading. For more information about calibration, see Chapter
4, “
Calibration
”. In dual or auto range modes, “HIGH” or “LOW” is
displayed to indicate the calibration of the high or low coefficient.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “
Calibration
”.
●
In the Main Menu, choose Calibration >
Cal CO2 Coefficient
.
Calibrate CO
2
Background
Calibrate CO
2
Coefficient
CALIBRATION:
>CAL BACKGROUND
CAL COEF
ZERO/SPAN CHECK
CAL O2 BACKGROUND
CAL O2 COEFFICIENT
RANGE AVG DIAGS ALARM
CALIBRATION:
>CAL BACKGROUND
CAL HI COEF
CAL LO COEF
ZERO/SPAN CHECK
CAL O2 BACKGROUND
RANGE AVG DIAGS ALARM