C-Link Protocol Commands
Calibration
Thermo Fisher Scientific
Model 410
i
Instruction Manual
B-29
coef o2
This command reports the O
2
coefficient. The example below reports that
the O
2
coefficient is 1.000.
Send:
coef
o2
Receive:
coef o2 1.000
set coef o2
value
This command sets the O
2
coefficient to a user-defined
value
, where
value
is a floating-point representation of the coefficient. The example below sets
the O
2
coefficient to 1.005.
Send:
set coef o2 1.005
Receive:
set coef o2 1.005 ok
coef 0
coef 1
coef 2
high coef 0
high coef 1
high coef 2
low coef 0
low coef 1
low coef 2
The “coef 0”, coef 1”, and coef 2” commands report the coefficients of the
curve developed from the Hi Multi-Point Calibration in single range mode.
The “high coef 0”, high coef 1”, and high coef 2” commands report the
coefficients of the curve developed from the Hi Multi-Point Calibration in
dual or auto range mode. The “low coef 0”, low coef 1”, and low coef 2”
commands report the coefficients of the curve developed from the Lo
Multi-Point Calibration in dual and auto range mode. The following
example reports the coefficient 0 having a value of 1.005.
Send:
coef
0
Receive:
coef 0 1.005 ok
co2 bkg
This command reports the current CO
2
backgrounds. The following
example reports that the CO
2
background is 1.4 ppm.
Send:
co2
bkg
Receive:
co2 bkg 1.4 ppm
set co2 bkg
value
This command is used to set the CO
2
background to a user-defined
value
,
where
value
is a floating-point representation of the background in the
currently selected units. The following example sets the CO
2
background
to 1.4 ppm.