Servicing
Filter Wheel Replacement
7-14
Model 410
i
Instruction Manual
Thermo Fisher Scientific
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the motor and wheel assembly by removing the three motor
plate Allen screws holding the motor plate to the optical bench
Figure
7–7
or
Figure 7–8
as appropriate.
Figure 7–7.
Removing the Motor (Standard)
Figure 7–8.
Removing the Motor (High Level)
3.
Remove the cross recessed screw on the bottom of the motor plate.
Motor Plate Screws (3)
Motor Plate Screws (3)