Serial Interface
72
HAAKE Phoenix II P1
Version 3.07
4.
Start the program by entering the respective program no.
(This START command corresponds to the START of the
ramp via the menu function) without starting the start tem-
perature:
W
[
RS
[
<
program no.
>
For starting the start temperature the instruction
W
[
AS
[
<
program no.
>
must be input before
This will transfer the program from the PC to the circulator.
The PC can now be switched off or be used for other tasks.
The interface cable can be unplugged from the circulator.
After switching off the circulator the programs are lost.
Example of a program:
Program no. 3 with 6 segments and
23 program cycles
1.) W
[
RN
[
3
2.) W
[
SA
W
[
SN1
W
[
SP
[
1
[
180.0
[
190.0
[
300
[
0.1
3.) W
[
SA
W
[
SN2
W
[
SP
[
2
[
190.0
[
185.0
[
600
4.) W
[
SA
W
[
SN3
W
[
SP
[
3
[
185.0
[
205.0
[
240
5.
W
[
SA
W
[
SN5
W
[
SP
[
4
[
205.0
[
185.0
[
240
6.) W
[
SA
W
[
SN6
W
[
SP
[
5
[
185.0
[
205.0
[
240
7.) W
[
SA
W
[
SN7
W
[
SP
[
6
[
205.0
[
205.0
[
180
8.) W
[
RS
[
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com