Operating
41
HAAKE Phoenix II P1
Version 3.07
14.2 Setting the desired set value
14.2.1 Setting the desired set temperature
The settings apply to the internal or external control sensors
according to the selected mode Intern/Extern (see chap-
ter13.2 ).
14.2.2 Setting and selecting the set values
1
Select „Set values“ in the main menu.
⇒
In the left area of the display the 4 stored set tempera-
tures with the accompanying RTA values are displayed
S1--S4.
There are two possibilities:
•
Change set value
Select the set temperature you want to change by pres-
sing the assigned function key and then activate
„Change set value“.
Now you select the plus minus sign (+ or -- ) of the new
set value by pressing the corresponding function key.
Then you enter the new value via the number keys with
3 digits before and 2 digits after the decimal point.
After entering the last digit you are asked whether you
want to activate the value. If not, you can enter a set
temperature again, otherwise you get back to the set
values submenu.
In case of a wrong entry, all empty digits have to be
entered before you can enter a new value.
Aborting during the entry is not possible.
•
Taking over and activating existing set values
For this select the desired set temperature by pressing
the assigned function key and thereafter take over the
value by „Activate set value“.
After that, the main menu is displayed automatically.
⇒
After pressing „Start
IDENT
flashes in the display if the
identification is activated. The FuzzyStar-controller is
determining the right parameter. It can happen that
heating or cooling is interrupted during this process
and the displayed temperature changes. This is
because the control unit requires some time to deter-
mine the parameters.
To deactivate the identification, select ”Settings” and ”Con-
trol”. Switch ”Identification” to ”OFF”.
To switch off the identification please see chapter 14.5 .