12. Connecting Up
12.1 Connecting to the mains
Only attach this unit to mains sockets with a grounded earth.
Compare the local mains voltage with the specifications writ-
ten on the name plate. Voltage deviations of +/- 10% are per-
missible. The socket must be rated as suitable for the total
power consumption of the unit.
Make sure that the temperature controller Phoenix is safely
connected to the refrigerated unit with the cable
19
.
12.2 Checking the liquid circuit
Before switching on, check again to make sure that the pres-
sure and suction ports are either connected with each other
or blocked with covering plugs or alternatively if an external
object is to be temperature controlled, that the hoses are
connected correctly and secured (see chapter 9.).
12.3 Changing the mains plug (e.g. for Great Britain)
This should only be carried out by qualified spe-
cialist personnel!
The mains cable wires have the following colors:
Brown
=
Live
Blue
=
Neutral
Green/Yellow=
Earth
Note amperage on rating plate!