Operating
47
HAAKE Phoenix II P1
Version 3.07
temperature:
before and after running the program
while running the program:
set temperature
actual temperature
program segments:
run time
min
6. At this point the heating is inactivated. The temperature
decrease given in segment no. 4 is too fast for a circulator
without cooling aggregate. Thus the actual temperature
decrease is slower than the set temperature.
7. After segment no. 4 is finished no. 5 starts. At first the set
temperature is higher than the actual temperature
.
The
heating is activated, until the linear rising set temperature
equals the actual temperature.
8. The heating is activated again. As the temperature gradi-
ent set in segment no. 5 is too fast, at the end of the seg-
ment the set end temperature has not been reached yet.
In segment no. 6 the end temperature of no. 5 is to be
maintained. The unit heats as long as this temperature
has been reached
.
Then by means of circled heating the
temperature is maintained constant until the end of the
segment.
9. At the end of segment no. 6 the program has been fin-
ished. Now the unit maintains the last set temperature of
the program, i.e. 205
°
C.
The times of segments no. 3 and no. 4 should be
extended so that the circulator can keep the given
values.