Serial Interface
65
HAAKE Phoenix II P1
Version 3.07
Standard
Extended
NAMUR
Command
V
R_V1
--
Version
... of the operating software
B
--
--
R_BS
R_FB
R_FE
--
--
--
Operating status
call up fault messages (19.8)
call up fuzzy control
call up fuzzy error number
I or F1
F2
--
--
IN_PV_1
IN_PV_2
Actual temperature
Call up
actual temperature (internal)
actual temperature (external)
S
S__
<value>
R_SW
W_SW_
<value>
**)
IN_SP_1
OUT_SP_1___
<value>
Set temperature
(resolution 0,01
°
C)
Call up set temperature
Select set temperature
<value>
= {00000 ... 20000} => 0
°
C ... 200,00
°
C
*)
<value>
= {00000 ... --5000 } => 0
°
C ... --50,00
°
C
e.g.: set temperature= 20,0
°
C => ”S__02000”
set temperature= --10,5
°
C => ”S --1050”; ”--” replaces 1st digit
C
CE
C__
<o.s.><value>
CE_
<o.s.><value>
R_CI
R_CE
W_CI
<_value>
**)
W_CE_
<value>
**)
IN_SP_2
IN_SP_4
OUT_SP_2___
<value>
OUT_SP_4___
<value>
RTA (correction factor c)
(resolution 0,01
°
C)
Call up RTA (internal)
Call up RTA (external)
Set RTA (internal)
Set RTA (external)
<o.s.>
= {+,--}; operational sign
<value>
= {0000 ... +/--0255} => 0
°
C ... +/--2,55
°
C
”blank” and ”+” can be used as positive signs,
e.g.: c = 1,23
°
C => ”c__0123” or ”c_+0123”
HL
LL
--
--
R_HL
R_LL
W_HL_
<value>
**)
W_LL_
<value>
**)
IN_SP_6
IN_SP_7
--
--
Temperature limit values
(resolution 0,01
°
C)
Call up high limit temperature
Call up low limit temperature
Set high limit temperature
Set low limit temperature
AL
W_AL
OUT_MODE_4_0
OUT_MODE_4_1
Alarm
... triggering (main relay mis-
sing, heating and pump off)
Alarm confirming
” _ ” = blank
*)
When a new set temperature is set while the start segment is triggered or when the ramp
is set on pause, it is stopped and continues to work with the new set value.
**)
at values with decimal places, put in a point e.g. 20.01 (no comma!!!)