13.5 Setting Status
13.5.1 Version numbers
For servicing or in case of product enquiries the version
numbers of the software or the control should be given.
Select „Settings“ and then „Status“. The version numbers
will be shown.
13.5.2 Operation status
For information about the operating status (alarm circuit of
the instrument or cooling device (C-AL...), internal or exter-
nal control and interface) please select „Settings“ and then
„Status“.
“WT” displays the operating hours of the unit.
In addition, the set overtemperature value is displayed in this
menu point without changing the setting (see Section 15.).
13.6 Settings the pump speed
In order to adjust the circulator to the respective application,
the pump speed can be varied in three steps. In the „Set-
tings“ menu the pump speed can be selected by pressing the
pump function several times.
„Pump fast“ = (maximum power): This setting is only neces-
sary when external systems are connected or when the bath
volumes are very large to achieve better controlling results.
„Pump medium“ = (set by the factory): This speed is suffi-
cient in most cases, especially at bath applications.
„Pump standard“: This setting should only be selected when
great turbulences have to be avoided in the bath.
It is also recommendable when you want to work in the refrig-
erated bath at the lowest bath temperatures that can be
reached because the heat supply by the pump is very low at
this setting.