Serial Interface
69
HAAKE Phoenix II P1
Version 3.07
Extended
Command
Fault F126
= ”RAMP ERROR”
If you select a segment for which not all the parameters have been defined, error message
”F126” appears. In this case either select another segment or define the parameters.
W_AS_<no>
W_RS_<no>
Trigger start temperature and without start
ramp.
Immediate start of the ramp without triggering
the start temperature
W_RB
W_RP
Stop ramp
Interrupt ramp
You can interrupt the program with W_RP. Then the segment time is hold and the momenta-
nous temperature is maintained. The interrupted program can either be continued with W_RS
or another segment can be selected with W_RB.
Fault F127
= ”PAUSE ERROR”
The command W_RP ”interrupt ramp” is only available while a program is running, i.e. if the
commands W_AS_<no.> or W_RS_<no.> have been entered before.If this is not the case
”F127” appears when entering the command W_RP ”interrupt ramp”.
Fault F001
= ”COMMAND UNKNOWN ”
If a command is entered that is not defined in any set of commands, error message F001
appears. In this case enter one of the commands listed above.
The single ramps can be repeated cyclically:
R_RZ
Call up number of cycles
W_RZ_
<number>
Set number of cycles
Call up for information:
R_CR
R_CS
R_RN
R_SA
R_RI
Read current (active) ramp number.
Read current (active) ramp segment.
Read ramp number which is currently run.
Read number of segments of the current ramp.
Read remaining ramp time and internal sensor
value.
R_RE
Read remaining ramp time and external sensor
value.
R_XR
Read status ramp continuation
Answer „ XR_< program no. >_< segment no. >_< segment remain time >_
< setpoint step >_< actual ramp setpoint >_< segment end temperature > „
” _ ” = blank