Internal temp.
External temp
Set value
°
C
°
C
°
C
22.68
21.20
40.00
17:00:00 17.11.00
ALARM
Cooling
Fault Displays
56
HAAKE Phoenix II P1
Version 3.07
16.7 Error in connection with cooling units
If the compressor of the cooling unit is overloaded the
circulator will be switched to fault status: ”COOLING”.
Allow the unit to cool down for a few minutes and then
try to start up again. If the fault occurs again ...
⇒
return unit for servicing.
Bath temperature < L-Limit
If the low temperature limit (L-Limit) is changed so that
is above the actual temperature, fault 15 occurs.
⇒
In the „Settings“ / „Temp. settings“ / „Limits“ menu
select a low temperature limit that is below the
actual temperature.
⇒
ALARM is still visible on the display.
⇒
The reset key must be pressed in order to start up
the unit again. ALARM disappears on the display
and the unit can be operated again.