Start/
No
End/
Time
Back
36 : Ramp functions view/selection
Var./K
°
C
°
C
1
20.00
70.00
0.10
--
00:30:00
R
S
Edit
Ramp 1/1
t/hh:mm:ss
Back
42 : Graph ramp
070.00
°
C
00:00:00
020.00
045.00
00:30:00
Operating
44
HAAKE Phoenix II P1
Version 3.07
The segment you want to edit can be selected with the two
arrow keys above and under the „S“ at the right of the display
and can be changed with „Edit“.
After having ended the programming you go back to the
ramp menu via „Back“.
14.4.2 Viewing stored programs
Via the function „View/ Selection“ the temperature programs
which have been stored can be viewed and selected.
The display in this menu corresponds to the settings menu
as described in 14.4.1.
Then the program segments are displayed in the table. With
the two arrow keys above and under the „S“ the desired seg-
ments can be selected. When required, they can be changed
with „Edit“. After selecting and/or changing the temperature
program you get back to the ramp menu with „Back“.
14.4.3 Graphical display of the active temperature
program
With the function „Graphic“ of the ramp menu the course of
the set values of the temperature program is displayed
graphically and can be seen at one glance.
The x--axis shows the time and the y--axis the accompanying
set temperatures.
You get back to the ramp menu by pressing the „Back“ func-
tion key.
14.4.4 Deleting a program
The currently selected ramp program can be deleted quickly
by pressing the „Delete ramp“ function.
14.4.5 Selecting beep
With the function „Beep“ you can select whether a signal
sounds after the end of each segment, after the end of each
ramp or if there is no signal.