16. Fault Displays
An acoustic signal is sounded and ”ALARM” is shown on the
display. The heating element, the pump and if available, the
compressor are completely switched off.
16.1 Excess temperature
The excess temperature protection can be triggered if:
•
Excess temperature has been set too closely to the
desired working temperature
⇒
increase value slightly according to specifi-
cations made in chapter 15.1.1.
•
the control function is defective
⇒
Return unit for servicing.
16.2 Low liquid level cut-off
The low liquid level protection can be triggered if:
•
there is not enough liquid in the bath
⇒
check for leaks, top up if necessary,
⇒
fluid has evaporated, replenish liquid.
16.3 Pump or motor overloading
The motor or pump is blocked:
⇒
If the circulator switches off again after a short
time, return the unit for servicing!