![Teledyne ProEM Series System Manual Download Page 90](http://html1.mh-extra.com/html/teledyne/proem-series/proem-series_system-manual_1080561090.webp)
90
ProEM System Manual
Issue 3
6.1.1
EXT SYNC Trigger Input
The selected Timing Mode {Trigger Response} determines how the camera will respond
to an External Sync pulse that is input at the EXT SYNC connector on the rear of the
camera. See
Figure 6-1: Rear of ProEM Camera
Things to keep in mind when setting up the External Sync pulse input are:
•
Pulse Height: 0 to +3.3V logic levels (TTL-compatible).
•
Pulse Width (trigger edge frequency): The time between trigger edges.
•
EXT SYNC Connector Impedance: High impedance.
•
Trigger Edge {Trigger Determined By}: rising or falling edge must be indicated
on the Experiment Setup
►
Timing tab {Trigger expander}.
4
4
11
-0
126
_
004
6
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...