Making the Camera-Circulator Connections for a CoolCUBE
II
For liquid-cooled cameras, the CoolCUBE
II
circulator provides a vibration-free method
of heat removal.
Perform the following procedure to connect the ProEM to a CoolCUBE
II
:
1. Make sure the camera and the circulator power switches are turned off.
2. Make sure the circulator is 6 inches [150 mm] or more below the camera. The
vertical distance should not exceed 10 feet [3 m.] Typically, the camera is at table
height and the circulator is on the floor.
3. Make the coolant connections between the circulator and the camera. It does not
matter which hose from the circulator is plugged into a coolant port on the camera.
See
Figure 4-5: Typical CoolCUBE
II
Circulator
4. It is recommended that hoses be secured to the camera hose barbs with the clamp
supplied.
NOTE:
Make sure that there are no kinks in the hoses that impede
the coolant flow. Lack of sufficient flow can seriously harm
the camera and any resulting damage is not covered under
warranty.
Damage caused by water leaking into the ProEM voids the
warranty.
5. Unscrew the reservoir cap on top of the CoolCUBE
II
and make sure that the coolant
reservoir contains coolant. See
Coolant Ports
Reservoir Cap
4
4
11-
01
26
_0
020
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...