Chapter 7
Tips
117
7.3.2
WinX/32 Step-by-Step Procedure
The following procedure uses a ProEM:1600
(2)
camera. This same procedure can be
used to increase the spectral readout rate for a ProEM:1600
(4)
.
NOTE:
The Readout Times you get may differ from those reported
in this procedure. These values are intended to show the
effect of the changes on Readout Time reduction.
1. On the Experiment Setup
►
ADC tab, select:
•
Rate: 4 MHz
•
Readout Port: Multiplication Gain
2. On the Hardware Setup
►
Controller/Camera tab, select:
•
Vertical Shift = 2
s
3. Select Easy Bin from the Acquisition menu.
•
Select ALL. This will result in Full Vertical Binning.
See
Figure 7-2: Typical WinX/32 Easy Bin Dialog
4. Acquire a spectrum.
4
4
11
-01
26
_
007
2
Summary of Contents for ProEM Series
Page 1: ...www princetoninstruments com ProEM System Manual 4411 0126 Issue 3 September 26 2019 ...
Page 10: ...10 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 16: ...16 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 30: ...30 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 50: ...50 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 88: ...88 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 114: ...114 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 122: ...122 ProEM System Manual Issue 3 Figure 7 8 LightField Settings 4411 0126_0078 ...
Page 136: ...136 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 146: ...146 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 152: ...152 ProEM System Manual Issue 3 This page is intentionally blank ...
Page 161: ...This page is intentionally blank ...