Functional Overview
WFM 601i Serial Component Monitor
2–23
1. To set horizontal gain, push the CONFIG menu button. Select CALIBRATE,
then CAL SIG ON.
Select GAIN CAL ON
.
a. In 1 LINE SWEEP
(5
µ
s/div) there should be one-half cycle per major
graticule division; in 2 LINE SWEEP
(10
µ
s/div) there should be one
full cycle per major division.
b. If the RESET H CAL readout is on, push the adjacent bezel button.
c.
Pushing RESET H CAL reinstates the gain setting established at the last
calibration.
d. Check to see that there is one half cycle/division (1-line sweep) or one
cycle/division (2-line sweep).